| Symptom | Cause |
Action |
| Liquid drops out from the tip, pipetting volume incorrect |
Tip is loose | push tip on tightly |
| The shaft nut is loose | tighten the shaft nut | |
| The sealing surface damaged | replace damaged parts | |
| The piston damaged or seal worn out due to extensive use |
replace damaged parts then wash the housing and lubricate the seal and o-ring with silicone lubricant |
|
| The tip is damaged or worn out due to extensive use |
replace the tip with a new one | |
| Droplets of air appear in the aspirated liquid |
The pipette tip immersion is too shallow | immerse the tip deeper accordingly to the instruction |
| Liquid is aspirated too fast |
decrease the speed of aspiration | |
| The pipette tip is inadequately pressed onto the pipette cone |
press the tip harder |
|
| Droplets of liquid remain in the pipette tip |
The pipette is emptied too fast |
decrease the speed of pressing the pipette pushbutton |
| The tip wettability has increased due to extensive use |
replace the tip with a new one |
Methods of solving problems and complete pipette assembly are described in detail in each instruction manual.